The digital patient
Abstract
Despite efforts, the detection of patients who are deteriorating in hospital is often later than it should be. Several technologies could provide the basis of a solution. Recording of vital signs could be improved by both automated transmission of the measured parameters to an electronic patient record and the use of unobtrusive wearable monitors that track the patient’s physiology continuously. Electronic charting systems could make the recorded vital signs readily available for further processing. Software algorithms could identify such patients with greater sensitivity and specificity than the existing, paper-based track-and-trigger systems. Electronic storage of vital signs also makes intelligent alerting and remote patient surveillance possible. However, the potential of these technologies depends strongly on implementation, with poor-quality deployment likely to worsen patient care.
Introduction
Over the past decade, there has been a considerable focus on the identification and response to unexpected clinical deterioration occurring in hospitalised patients. Much of this work has been prompted by studies demonstrating that many deteriorations are not detected in a timely fashion, leading to unplanned admission to the intensive care unit or to cardiac arrest, which would have been avoidable had appropriate care been instituted at an earlier stage.1 Factors contributing to avoidable cardiac arrests have been examined in several studies,2,3 most notably the National Confidential Enquiry into Patient Outcome and Death (NCEPOD) report in 2005.4 These studies identified a wide variety of systemic problems, ranging from failure to record patient vital signs with sufficient frequency to failure to respond appropriately to a patient identified as having deteriorated. Following National Institute for Health and Clinical Excellence (NICE) guidance,5 hospitals in the UK have implemented early warning scores (also called ‘track-and-trigger’ scores) to aid detection of the deteriorating patient on the general ward. Yet, despite these scores, late detection of such patients remains a problem.6
As part of its ongoing work to address this issue, the Royal College of Physicians recently published proposals for a National Early Warning Score (NEWS),7 advocating that it should be widely adopted across National Health Service (NHS) hospitals with the aspiration that improvements in patient care will occur through standardisation of alerting thresholds and escalation procedures. Implementation of NEWS will require hospitals to change observations charts, adjust escalation pathways and retrain staff. Therefore, this represents an opportunity to consider how recent technological developments could improve the identification of patients who are deteriorating.
As shown in Fig 1, systems for identification and treatment of patients who are deteriorating may be divided into the following components: devices to measure markers of abnormality (patient monitors recording vital signs); systems to record the collected data (often a paper chart); systems to analyse the recorded data (such as an Early Warning Scoring System); and systems to respond to the results of the analysis (the alerting and escalation pathways of a hospital). In this article, we provide an overview of the technological options for each component and how they might develop in the future. Although there is overlap with technologies used in intensive care environments, we focus primarily on systems applicable to Level 1 wards.
Measurement of vital signs
The existing standard of care in British hospitals is continuous monitoring in high dependency and intensive care areas and intermittent monitoring using ‘spot-check’ monitors on level 1 wards. In exceptional cases, the latter can be augmented by continuous monitoring using either portable bedside monitors or wireless telemetry devices on wards where wireless access points are installed.
The typical hardware used for recording vital signs on the level 1 ward can be improved in several ways. Introduction of spot-check monitors capable of wirelessly transmitting the recorded observations to an electronic charting system should lead to process improvements by eliminating transcription errors and speeding the process of recording observations. Wireless spot-check monitors can either be purchased as systems with integrated wireless networking cards or created by retrofitting existing monitors with an additional device to enable wireless transmission.8
An alternative strategy for improving the observation recording process is more widespread use of continuous (or near-continuous) recording of vital signs using multiparameter monitors, with any intermittent measurements, such as measurement of blood pressure, being triggered by the monitor. Theoretically, this should lead to detection of abnormal physiology at the earliest possible moment and reduce the incidence of observations being missed or being taken too infrequently. However, when this approach was trialled using existing bedside monitors in both medical and surgical populations, no benefit was found on intention-to-treat analysis.9 Strikingly, only 16% of patients assigned to automatic monitoring remained connected to their monitors for the entire 72 hour period mandated by the study protocol. The most common reasons cited for removal of monitoring were patient request and to enable patients to mobilise, clearly demonstrating the ergonomic inadequacy of conventional bedside monitors for long-term recordings in ambulatory patients. For continuous monitoring to provide benefit in a general ward population, equipment needs to be more comfortable and less restrictive.
In response to this challenge, several manufacturers have developed novel monitors. Wearable cardiac telemetry devices are already established in the hospital environment. However, most devices designed for in-hospital use rely on a radio-frequency network specific to the device to transmit the acquired data. Installing the requisite network access points is costly and ties the hospital to a single provider of telemetry. The devices themselves are bulky compared with some of the newer monitors under development (see below) and most are restricted to monitoring echocardiography (ECG), although a few can also record oxygen saturation. Such solutions do not remove the need for intermittent monitoring, because the other vital signs also need to be recorded.
The newer monitors on the market (examples of which are listed in Table 1) are generally closer to being wearable for prolonged periods and many record a more diverse set of physiological parameters, including motion (which can be used both to detect patient falls and to measure patient activity levels). Only one device available to date, the Sotera Wireless VisiMobile, is explicitly targeted at the hospital environment and claims to measure all five commonly recorded vital signs. The remainder are either designed to cater for outpatient monitoring or for short-term use in high-risk environments.
Unfortunately, these newer, wearable monitors are not yet suitable for widespread deployment as clinical monitoring devices in a hospital environment, partly because of lack of suitable supporting software optimised for use by clinical staff in a hospital environment and partly because of the design choices made when targeting them for out-of-hospital use. In a recent assessment of three novel monitors, none managed to record data consistently for 24 h.10 However, given the considerable amount of interest in home monitoring using wearable monitors (the Department of Health spent over £30 million funding the recent Whole System Demonstrator trial11 to assess the benefits of telehealth), it is likely that these systems will mature rapidly.
Wearable sensors are not the only monitoring technologies being explored. Sensors woven into clothing,12 printed on transfer tattoos13 and embedded into furniture14 are also being developed. Monitors need not be physically attached to patients in any way: non-contact sensing technologies include video cameras,15,16 Doppler radar17 and ultra-wideband radar.18 Although such sensors are not currently sufficiently developed to enable mainstream use in a hospital environment, they do offer the promise of ubiquitous monitoring without impairment of patient mobility.
However, an ergonomic monitor that simply transmits data to a remote screen or stores the information locally is less useful than a wearable monitor connected to a system that stores the recorded data and makes it available for further analysis. The nature and potential benefits of analysis are discussed below. Such processing is greatly facilitated by the adoption of open standards for the storage and transmission of data. This has been widely recognised and consortia, such as the Continua Healthcare Alliance and Integrating the Healthcare Enterprise, have been formed by key industry players to promote the use of open standards. Their efforts are complemented by work done in the open-source community, a notable example being the OpenEHR project, whose standards for the storage, retrieval and exchange of patient data have been adopted by the NHS through the Connecting for Health program.
Charting of vital signs
Existing practice in most UK hospitals is to record the vital signs on a paper chart, with the nurse recording the observations, adding the scores for each vital sign parameter and then entering the overall early warning score on the chart. This approach has a risk of error in both assigning the correct score to each vital sign parameter and calculating the total score.19–21 Other well-described problems with the paper chart are poor legibility, inaccurate plotting and the possibility of misplacing or losing the chart. These flaws are frequently cited as reasons for the introduction of electronic charting systems.22
Such systems can be broadly divided into three categories: charting modules, which are part of an electronic patient record application; stand-alone charting systems; and charting systems that are designed to work solely with the wireless spot-check monitors described above.
Studies investigating the benefits of electronic systems report highly variable results, in part because their success or failure is defined in narrow terms, such as the speed of data entry or user acceptance,23 rather than from a broader clinical perspective (which in itself is only one of many pertinent perspectives24) and also because benefits measured in a non-clinical setting19 might not correspond to benefits on the ward. Furthermore, the introduction of electronic charting systems might have unintended consequences by paradoxically making data less easily accessible to nursing staff.25 Despite over two decades of development in this field, there are still many lessons to be learnt regarding the optimal design and functionality of such systems. Nevertheless, such systems do enable a transformative change in the way that vital sign data are used: storing the data electronically makes them available for more complex and varied processing than is afforded by calculation of an integer early warning score.
Processing of vital signs
Existing early warning scores are designed to be simplistic so as to facilitate calculation using pen and paper. This limits the sensitivity and specificity that they can hope to achieve. A completely electronic system enables integration of non-vital sign parameters (such as laboratory results) and patient-related factors (such as age and comorbidities) into a scoring system,26 as well as facilitating complex analysis of vital signs. Rather than assuming that vital signs are independent variables, the subtle interactions between them can be assessed in a probabilistic manner to determine earlier whether a patient is clinically deteriorating. An example of such a system has been trialled on a surgical step-down unit in the USA27 and was found to be effective in predicting physiological instability before it became overt.
Trends of the vital signs over time28,29 or statistical derivatives30 can also be used to improve alerting algorithms. Alerting thresholds can be personalised on a patient-by-patient basis using short-term trends31 or historical steady-state values. With continuous data, as might be provided by wearable monitors, it is possible to use the recorded waveforms to derive parameters with prognostic significance, such as heart-rate variability,32 ECG dispersion33 or the entropy of the signal.34
In the future, it is likely that computerised algorithms will be used to augment or replace current early warning scores. It might be that, rather than one algorithm reporting a patient’s status, multiple complementary algorithms will be used simultaneously, each reporting on the patient’s condition from a different perspective. This should resolve the tension between providing a probability of imminent death (NEWS and its precursor ViEWS)35 and providing early warning of abnormal physiology to enable timely intervention.36,37
Alerting
Electronic acquisition and storage of observations provides further potential for benefit through the generation of ‘smart alerts’. The most commonly used alerting technology on the general ward today is the local single parameter alert, whereby any one of the vital signs breaching a pre-set threshold causes the bedside monitor to sound an audible alarm. Unfortunately, this approach usually leads to a large number of false alarms, partly because the thresholds are often left at default settings, irrespective of the condition of the patient who is being monitored, and partly because signal artefact is often misinterpreted by the monitor. One study in an operating theatre environment showed that 75% of alarms were spurious and only 3% indicated serious patient risk.38 In an environment where patients move freely, the false alarm rate can be even higher. The problem is exacerbated because the audible alarms from patient monitors are just one group of a host of audible alarms emitted from ward devices, ranging from inflatable mattresses to patient call bells. This quickly leads to so-called ‘alarm fatigue’, which causes staff to cease responding to alarms in a timely fashion. In the most extreme cases, this can lead to missing the fact that a monitored patient is dying.39
Intelligent alerting systems might help to address this problem. These might either generate alarms based on multiple parameters, in a similar fashion to an early warning score, or suppress alarms until a single parameter has consistently exceeded a pre-set threshold, thereby avoiding the problem of alarming owing to transient episodes of artefact.
Computer-generated alarms need not sound at the bedside; they can be delivered remotely to a nursing station, as happens in many level 2 and level 3 areas, or they can be linked directly to the clinicians’ pagers. One study on an orthopaedic ward found that sending alerts from pulse oximeters directly to medical emergency teams led to a significant reduction in number of intensive care unit admissions, despite the fact that the thresholds for alerting were set to an arterial oxygen saturation of less than 80% and a heart rate of under 50 beats per minute or over 140 beats per minute.40 Alerting can be based on the hospital early warning score and the same abnormal score can be used to generate multiple alerts that are sent in a sequential fashion to different clinical team members depending on the severity of the abnormality and the response times. A study of one such system suggested that this leads to clinical benefit, although the results were not definitive.41 Such systems might help avoid the problem of ward staff recognising that a patient is deteriorating but not calling for help owing to sociocultural factors.42
Other approaches to remote alerting involve setting up a centralised monitoring room where patients attached to telemetry devices are surveyed by monitor technicians who call the ward staff via a dedicated hotline if they identify important changes in a patient’s condition.43 Remote surveillance of patients could also be used to identify patients before they are escalated44 or as part of a handover process, such as occurs to a hospital-at-night team.
Potential for benefit
The concepts outlined in this article have been discussed in the literature for over two decades. During that time, significant progress has been made in the implementation of these ideas. Further refinement is especially needed in algorithms for detecting patients who are deteriorating, along with better methods for assessing their merits. Endpoints that have been previously used, such as length of stay, intensive care unit transfer and mortality, are influenced by a range of factors that can lead to erroneous characterisation of the effectiveness of the systems under study37 and make it difficult to make rational choices about the best algorithms to use.7
Realisation of benefit is strongly dependent on the quality of the implementation,45 user acceptance and the cultural and organisational factors surrounding the introduction of these systems.46 Poor implementation not only fails to result in benefit, but can also make care worse.47–50
Considerations surrounding the implementation of a technological solution should not be restricted to the technology and the infrastructure that directly supports it. Each technology is supported by a plethora of other systems and the impact on each of these must be assessed. For instance, according to the report that describes it, NEWS will result in an increased alert rate in many hospitals if the suggested triggers for escalation are used. One estimate, based on a database of vital signs from a mixed medical and surgical population, is that in a 1,000-bed hospital, 500 ‘amber’ alerts and 200 ‘red’ alerts would be generated per day.51 This might necessitate reorganisation of existing response teams or deployment of extra resources to ensure that the alerts are addressed in an effective and timely manner. Organisations that introduce NEWS without providing capable response mechanisms risk the alerts being ignored on occasion, with all the problems that this entails.
Care must also be exercised in identifying the assumptions made when justifying the introduction of new technology. There is yet to be any conclusive evidence that early warning scores improve patient outcomes,52 or that minimising errors in recording observations or calculating early warning scores has any significant clinical impact. That these process changes will be beneficial might seem to be self-evident until one considers that nurses frequently rely on cues other than the observations chart to determine whether a patient is deteriorating53 and use the observations to support their clinical intuition. Introduction of poorly designed systems can discourage nurses from utilising their clinical impressions because there is nowhere to record them or clinical judgement is removed from the escalation process.25
Conclusion
There are a vast array of technologies, currently available and in development, that promise to improve patient care significantly. Those with the greatest potential to bring about improvement are the systems that electronically store patients’ observations and make them available for further processing. Such systems offer benefits both directly, by addressing flaws in the recording process, and also by acting as a foundation upon which other safety systems, such as computerised algorithms and remote alerting systems, can be deployed. For full realisation of the clinical benefits, great care must be taken in the implementation of these technologies, because poor-quality deployment could lead to worse patient care.
- Royal College of Physicians
References
- ↵
- Vincent C,
- Neale G,
- Woloshynowych M
- ↵
- ↵
- ↵
- Cullinane M,
- Findlay GP,
- Hargraves C,
- Lucas S
- ↵National Institute of Clinical Excellence. Acutely ill patients in hospital. London: NICE, 2007.
- ↵
- Findlay GP,
- Shotton H,
- Kelly K,
- Mason M
- ↵Royal College of Physicians. National early warning score (NEWS). London: Royal College of Physicians; 2012
- ↵
- ↵
- ↵
- Bonnici T,
- Orphanidou C,
- Vallance D,
- et al
- ↵
- Steventon AA,
- Bardsley MM,
- Billings JJ,
- et al
- ↵
- ↵
- Kim D-H,
- Lu N,
- Ma R,
- et al
- ↵
- ↵
- Wu H-Y,
- Rubinstein M,
- Shih E,
- et al
- ↵
- ↵
- ↵
- ↵
- ↵
- ↵
- ↵
- Van Der Meijden MJ,
- Tange HJ,
- Troost JJ,
- Hasman AA
- ↵
- ↵
- Mackintosh N,
- Rainey H,
- Sandall J
- ↵
- ↵
- ↵
- ↵
- ↵
- Hug CW
- ↵
- ↵
- ↵
- ↵
- ↵
- ↵
- Morgan RJM,
- Wright MM
- ↵
- ↵
- Block FEF,
- Schaaf CC
- ↵
- Wallis L
- ↵
- ↵
- ↵
- Shearer B,
- Marshall S,
- Buist MD,
- et al
- ↵
- ↵
- ↵
- ↵
- Delone WH,
- McLean ER
- ↵
- Han YY,
- Carcillo JA,
- Venkataraman ST,
- et al
- Stevenson JE,
- Nilsson GC,
- Petersson GI,
- Johansson PE
- ↵
- ↵
- Greengross P,
- Beaumont K,
- Killen R
- ↵
- ↵
- Intelesens Aingeal, 2013. www.intelesens.com/hospitalmonitoring/aingeal.html [Accessed 4 April 2013].
- Nonin WristOx2 3150, 2013. www.nonin.com/PulseOximetry/Wrist-Worn/WristOx2-Model-3150 [Accessed 4 April 2013].
Article Tools
Citation Manager Formats
Jump to section
Related Articles
- No related articles found.
Cited By...
- Protocol for a systematic review assessing ambulatory vital sign monitoring impact on deterioration detection and related clinical outcomes in hospitalised patients
- Reliability of a wearable wireless patch for continuous remote monitoring of vital signs in patients recovering from major surgery: a clinical validation study from the TRaCINg trial