Table 2.

Agreement between the different locations and distances of the IR temperature measurement devices and between the SpotOn and other IR temperature measurement devices

IR deviceMean (SD)Mean difference (LOA)aICC (95% CI)
IR gun
Location: eyes, nose and lips0.15 (0.04, 0.27)
Distance head: 3, 6 and 9 cm0.71 (0.63, 0.78)
Distance temple: 3, 6 and 9 cm0.72 (0.65, 0.79)
Tympanic thermometer
Left and right ear0.66 (0.54, 0.75)
SpotOn versus IR device
SpotOn36.9 (0.37)
IR gun location
Eyes36.9 (0.36)0.07 (–0.88, 1.02)0.13 (0.00, 0.30)
Nose35.5 (1.26)1.40 (–1.15, 3.94)0.00 (0.00, 0.20)
Lips36.5 (0.38)0.41 (–0.49, 1.31)0.25 (0.08, 0.41)
IR gun distance
Forehead: 3 cm36.7 (0.30)0.19 (–0.58, 0.97)0.32 (0.14, 0.47)
Forehead: 6 cm36.6 (0.22)0.35 (–0.43, 1.14)0.15 (0.00, 0.32)
Forehead: 9 cm36.4 (0.23)0.51 (–0.32, 1.35)0.07 (0.00, 0.24)
Temple: 3 cm36.7 (0.48)0.24 (–0.93, 1.41)0.05 (0.00, 0.22)
Temple: 6 cm36.5 (0.42)0.42 (–0.65, 1.49)0.05 (0.00, 0.23)
Temple: 9 cm36.4 (0.46)0.57 (–0.60, 1.73)0.00 (0.00, 0.18)
Tympanic
Left36.3 (0.49)0.59 (–0.37, 1.55)0.36 (0.19, 0.51)
Right36.3 (0.49)0.64 (–0.31, 1.60)0.37 (0.20, 0.51)
IR image
Coronal plane36.4 (0.53)0.50 (–0.58, 1.59)0.27 (0.10, 0.43)
  • aSpotOn minus comparator measurement.

  • CI = confidence interval; ICC = intra-class correlation coefficient; IR = infrared.